Medical work is less about exotic tolerance and more about evidence: the right material grade, a surface that can be cleaned, edges that will not cut a glove, and a record that survives an audit two years later.

Stainless and titaniumBurr-free edgesBatch traceability
01

Part families

Surgical instrument handles and partsStainless, feel

Handles, forceps bodies and instrument shafts in 17-4 PH, 420 or 440C. The critical features are flatness on the mating faces, pivot hole alignment, and an edge break that is comfortable in a gloved hand.

Device housings and enclosuresSealing and appearance

Aluminium or stainless enclosures with gasket grooves, display windows and connector cutouts, finished by anodizing or passivation. Cosmetic faces are agreed on the drawing rather than argued at incoming inspection.

Dental components and prostheticsSmall, precise

Implant-adjacent parts, abutment holders, model components and prosthetic hardware. Small features, tight position tolerances, and a surface finish that will not trap residue.

Ventilator and fluid-path partsCleanability

Manifolds, valve bodies and connectors with intersecting passages and O-ring grooves. Burr at a passage intersection is a contamination risk, so deburr is a process step with a check, not a courtesy.

Inspection fixtures and test rigsRepeatability

Go/no-go gauges, holding fixtures and microfluidic test blocks. These parts are the measurement system, so their tolerance is tighter than the part they check.

02

Material and finish logic

Medical material selection
RequirementMaterialFinishNote
Corrosion plus strength17-4 PH (630)PassivationH900 or H1025 condition
Cutting edges, wear440CPassivation, polishHardened after machining, ground after
Chemical resistance316LPassivation, electropolishLow carbon for welded assemblies
Lightweight housing6061-T6AnodizeNot for repeated autoclave cycles
Insulation and wearPEEK, PTFENoneStress relief between rough and finish
Biocompatible metalTi Gr2 / TC4PassivationMill cert per heat lot
What we do not claimWe machine to the drawing and supply material certificates, dimensional reports and process records. We do not hold ISO 13485 or a medical device registration, and a website badge cannot substitute for one. If your programme requires a certified quality system, that must be settled with your quality team before the first order.
03

Cleanability starts at the drawing

A part with a blind 1.5 mm groove, a sharp internal corner and a cross-drilled hole breaking into it cannot be cleaned reliably, whatever the surface finish. Residue collects in geometry, and geometry is a design decision.

Practical rules: give every internal corner a radius of at least a third of the groove width, avoid blind holes intersecting passages at acute angles, and specify Ra 0.8 µm or better on any surface contacting tissue, drug product or cleaning fluid.

Design checks before release
Edge break0.2–0.5 mm on handled edges, stated on the drawing
Internal radii≥ 1 mm, ideally ≥ 0.5 × feature width
Blind holesDepth ≤ 4 × diameter, or through-drilled
Surface finishRa 0.8 µm functional, Ra 0.4 µm fluid contact
MarkingLaser engraving, away from contact surfaces
04

Documentation we supply

Material certificatePer heat lot

Mill certificate with grade, heat number, chemistry and mechanical properties, matched to the batch. For 17-4 PH the condition is stated, because H900 and H1025 machine and perform differently.

Dimensional reportFirst article and per batch

The callouts that affect function, measured on a CMM with the drawing's datum scheme. Bubbled drawing numbers make it reviewable in minutes.

Compliance declarationsRoHS and REACH

Issued against the batch, covering the material and the finish chemistry. Useful when the finished device has its own declaration obligation.

05

Industry questions

Do you sign an NDA before seeing patient-adjacent designs?

Yes, and we recommend it. Send the NDA first, then the model. Design files stay with the programme and are not used for any other customer.

Can you machine implant-grade titanium?

Ti-6Al-4V (TC4) is available with mill certification. Machining is slower than aluminium and the quote reflects tool wear, so it is worth confirming the grade is genuinely required before specifying.

How are burrs controlled on small instrument features?

By process: sharp tooling, controlled final passes, a defined deburr step, and magnification inspection on critical edges. For batches, vibratory finishing followed by edge inspection.

Is electropolishing available?

Yes, coordinated with the finishing supplier and quoted with the part. It removes 5–20 µm, so any tolerance on a polished surface must account for that before machining.

Send the drawing with material and finish condition stated.

For medical work, tell us the cleaning cycle and the documentation level early. Both change the process and neither can be added cheaply after the first batch.

Request a quote